|
Time-Dependent FUSE Calibration Effects
A FUSE White Paper
|
Appendix A
Segment 2A Gain Sag
16 December 2002 through 3 February 2003
David Sahnow
March 3, 2003
From 16 December 2002 to 3 February 2003, the high voltage on detector
segment 2A was reduced by approximately 250 volts, to a lower-than-optimal value,
due to high voltage transients on that segment.
Effects
While the
voltage was lowered, pulse heights were much lower than normal, resulting in
significant walk effects for LiF LWRS data. An
example of the effect of the voltage change on the pulse height is shown in the
following figure. It shows the pulse height for two different observations of
the same object taken in the LWRS aperture; since the exposure times were
slightly different, they have been normalized before plotting. In early
December, the modal gain for both SiC and LiF LWRS apertures was above 10. After the voltage change,
the LiF LWRS modal gain dropped to 5, while SiC was approximately 9. Values
below 8 are most likely to show significant walk problems, although this
depends on the position on the segment.
The figure also shows that the gain
has not dropped so much that counts have been lost below the lower threshold,
so the flux calibration was not affected.

The figure
below shows the effect of this gain sag on stim lamp
images taken during this period. The stim lamps
illuminate the detector directly in a roughly uniform manner. Shadows of the
grid wires at the front of the detector are visible because of the way the
lamps illuminate the detector. Each of the plots in the figure shows a y
projection of the photons in the region of the LiF
LWRS aperture collected during a stim lamp exposure.
The x range shown is from pixel 10950 to 11980, but the effects are similar
across the entire segment. All are binned by 2 in the x direction in order to
improve the signal-to-noise ratio.

(a) shows an exposure taken in February 2001, well before the
high voltage was adjusted. The positions of the grid wire shadows are noted
with dashed lines. This was a ~15000 second exposure taken in HIST mode. The shadows
of the grid wires are clearly visible.
(b) shows a ~7500
second TTAG exposure with far fewer counts taken on 6 December 2002, shortly before the voltage was
changed. The shadows of the grid wires are again quite prominent, although
there has been a slight shift in the location of a few. This shift is a sign of
the gain sag, and is the reason the voltage was raised a few days later.
(c) was taken a few weeks after (b), and under essentially the
same conditions, except that the voltage is now at the lower value. There has
been a significant shift in the positions of the shadows. These shifts are
typically 20 to 30 pixels.
(d) shows the same exposure as in (c), but this time with the
walk correction applied. The positions of the lines have shifted back towards
their correct locations, but the correction is not perfect; In addition, the
line widths have increased.
The next figure
shows the effects on a calibrated spectrum. The same calibration target was
observed both before and during the time when the voltage was low. The spectra
from these two HIST exposures show very clearly how the spectrum on segment 2A
is distorted because of the gain sag. The differences between the two are so
significant that it is difficult to tell that these are spectra of the same
object.

Conclusions
During the
period when the voltage was low on segment 2A, there is a significant shift in
the reported positions of many photons. For TTAG data, much (but not all) of
this is corrected by the walk correction module of the CalFUSE
pipeline. For HIST data, however, only a crude correction is applied, so data are
likely to be seriously affected. Line locations and equivalent width
measurements are likely to be in error on this segment, so results should be
carefully compared to data on other segments.
- Questions?
-
Please address questions to
fuse_support@pha.jhu.edu.
Last modified Monday, March 3, 2003
|